2. Spectral Data Collection ============================ 2.1 Which FT-MIR to use: ------------------------ Use any FT-MIR spectrometer connected to an **attenuated total reflectance (ATR)** accessory. 2.2 Parameters ---------------- This web application can be used for any of these data collection parameters (either 1 or 2): +-----------------------------------------+--------------+ | | | +=========================================+==============+ | Data Acquisition Mode | Absorbance | +-----------------------------------------+--------------+ | Number of Scans | 100 | +-----------------------------------------+--------------+ | Number of Background Scans | 200 | +-----------------------------------------+--------------+ | Apodization | Happ–Genzel | +-----------------------------------------+--------------+ | Resolution (cm⁻¹) | 2 | +-----------------------------------------+--------------+ | Zero-Fill Factor (ZFF) | 2 | +-----------------------------------------+--------------+ | Background validity time (minutes) | 30 | +-----------------------------------------+--------------+ 2.3 Cleaning ------------------ Clean the diamond ATR with **methanol** before initiating background scans and in between sample scans. 2.4 Contact ------------------ Ensure the flour sample is sufficiently in contact with the diamond ATR before data collection. 2.5 Number of Samples Needed ----------------------------- Flour from one or two seeds is sufficient to collect the spectral details from one representative sample.